Products PRODUCTS
Product name:PILATUS3 R CdTe for Laboratory
brand:Dectris
Place of Origin:Switzerland
The PILATUS3 R CdTe X-ray detector series combines the advantages of noise-free single-photon counting with the benefits of direct detection of hard X-rays in large cadmium telluride (CdTe) sensors. Direct detection in a hybrid pixel results in sharper signals and better spatial resolution than scintillator-based detectors. The CdTe X-ray detectors also provides almost 100% efficiency for Mo, Ag, and In radiation. Combined with noise-free single-photon counting, these capabilities permit DECTRIS CdTe X-ray detectors to achieve new levels of sensitivity and accuracy in the laboratory.
The PILATUS3 R CdTe X-ray detector series combines the advantages of noise-free single-photon counting with the benefits of direct detection of hard X-rays in large cadmium telluride (CdTe) sensors. Direct detection in a hybrid pixel results in sharper signals and better spatial resolution than scintillator-based detectors. The CdTe X-ray detectors also provides almost 100% efficiency for Mo, Ag, and In radiation. Combined with noise-free single-photon counting, these capabilities permit DECTRIS CdTe X-ray detectors to achieve new levels of sensitivity and accuracy in the laboratory.
Laboratory X-ray sources, particularly for hard X-rays, are much weaker than at the synchrotron and longer exposure times are required. Thanks to the absence of dark current and readout noise, PILATUS3 R detectors outperform all other technologies. Single-photon counting is also free of other sources of noise and inaccuracy intrinsic to charge-integrating CMOS detectors. DECTRIS Instant Retrigger technology in combination with an accurate count rate correction have enabled advanced synchrotron high-energy experiments and are hence fully compatible with all laboratory applications.
The PILATUS3 R CdTe detector series offers high-efficiency direct detection and noise-free single-photon counting — a winning combination for your laboratory.
技术参数:
PILATUS3 R CdTe |
1M |
300K |
300K-W |
100K |
探测器模块数量 |
2 × 5 |
1 × 3 |
3 × 1 |
1 × 1 |
有效面积:宽×高 [mm²] |
168.7 x 179.4 |
83.8 × 106.5 |
253.7 x 33.5 |
83.8 x 33.5 |
像素大小 [μm²] |
172 x 172 |
|||
总像素数量 |
981 x 1043 |
487 × 619 |
1475 × 195 |
487 x 195 |
间隙宽度, 水平/垂直(像素) *每个模块之间水平间隙增加1个像素 |
7* / 17 |
- */ 17 |
7* / - |
-* / - |
非灵敏区 [ % ] |
7.8 |
5.7 |
1.1 |
0.2 |
缺陷像素 |
< 0.1% |
|||
最大帧频 [Hz] |
5 |
20 |
20 |
20 |
读出时间 [ms] |
7 |
|||
点扩散函数 |
1 pixel(FWHM) |
|||
计数器深度 |
20 bits(1,048,576 counts) |
|||
功耗 [W] |
165 |
30 |
30 |
30 |
尺寸(WHD)[mm³] |
265 x 286 x 455 |
158 x 193 x 262 |
280 x 62 x 296 |
114 x 69 x 118 (探头) |
重量 [kg] |
25 |
7.5 |
7.0 |
0.9(探头) |
模块冷却 |
水冷 |
水冷 |
水冷 |
水冷 |
外接触发电压 |
5V TTL |
5V TTL |
\ |
\ |
PILATUS3 R CdTe 1M
PILATUS3 R CdTe 300K
PILATUS3 R CdTe 300K-W
PILATUS3 R CdTe 100K